We don't just swap the part that failed. We rebuild every part that wears out, so the device runs quietly for years.

Compressor repair

The compressor is the heart of the concentrator. We open it, clean it and rebuild it with new consumables.

  1. 1Replacement of the cup seals (the rings on the pistons)
  2. 2Cleaning the pistons of residue from use
  3. 3Honing and cleaning the cylinders
  4. 4Honing and cleaning the compression heads
  5. 5Replacement of the reed valves
  6. 6Bearing replacement where needed — only the part is charged, not the labour
  7. 7New O-rings throughout, especially at the top of the pistons
  8. 8Final check of pressure and air volume at the outlet
  9. 9Pressure relief valve test with the outlet restricted

Valve repair (pilot valve)

If the valve is not renewed, oxygen purity drops quickly and the zeolite is put at risk.

  1. 1Replacement of all elastomers and sealing O-rings
  2. 2On the Respironics EverFlo the valve is replaced with a new one
  3. 3Leak test under pressure before it is connected to the zeolite sieve beds

Zeolite sieve bed refill

Done in a moisture-protected room, so the sensitive zeolite lasts longer.

  1. 1Opening and emptying the sieve beds
  2. 2New filters at the top and bottom of each bed
  3. 3Branded Lithium or Activated Alumina zeolite, depending on the model

Checks at no extra cost

Every refurbishment also includes the following checks.

  1. 1Checking for and sealing leaks in the air and oxygen circuit
  2. 2Checking the run capacitor's capacitance and replacing it where needed
  3. 3Setting the outlet pressure to the value in each model's service manual
  4. 4Internal dust cleaning with 10 bar air
  5. 5Pressure regulator check
  6. 6Circuit board repair

How the repair works

  1. 1

    You let us know

    Fill in the repair request or call us. Always note the serial numbers.

  2. 2

    You send the device

    Through our partner carrier in your city. We pay for shipping to Chania.

  3. 3

    We tell you the cost

    We inspect the device and tell you the work needed and the final amount before we start.

  4. 4

    Refurbishment and measurement

    We record oxygen purity at 2, 4 and 5 litres before the device leaves.

  5. 5

    Return with warranty

    After payment, the device comes back with a written warranty and an itemised list of work.